Line 58: Line 58:
  
 
In order for the SiPMs to operate, they require a bias voltage of 56 V (with the first signals visible at 53 V and no exceeding 57 V). We use a dedicated power supply to provide this bias. The bias supply and the signal cables must be fed through a feedthrough on the side of the chamber into the vacuum. The schematic created for the cabling through the three feedthroughs in place is shown below. This also includes the cables for the temperature readouts. This feedthrough has been creating a significant level of noise on the readouts. To reduce some of this noise, a Ferrite ring has been wound (twice) around the connecting cable. The connections of the cables to the inside of the feedthrough are currently individual cases and are very difficult to manually connect – a better solution to this will be considered.
 
In order for the SiPMs to operate, they require a bias voltage of 56 V (with the first signals visible at 53 V and no exceeding 57 V). We use a dedicated power supply to provide this bias. The bias supply and the signal cables must be fed through a feedthrough on the side of the chamber into the vacuum. The schematic created for the cabling through the three feedthroughs in place is shown below. This also includes the cables for the temperature readouts. This feedthrough has been creating a significant level of noise on the readouts. To reduce some of this noise, a Ferrite ring has been wound (twice) around the connecting cable. The connections of the cables to the inside of the feedthrough are currently individual cases and are very difficult to manually connect – a better solution to this will be considered.
 
**Update 5/10/22**: A new connector has been built with permanent connection of the SiPM bias/signal wires to their correct feedthrough positions. This reduces the time required to insert the pins and prevents the chamber from becoming dirty and needing additional cleaning. The anode wire for the SiPM is connected to the wire marked with kapton tape and the cathode to the bare wire. This has been tested in vacuum and after cleaning with isopropanol and wipes, allows the vacuum to drop to 10^-3mbar in ~30 minutes, with an observable signal still readable.
 
  
 
[[File:Feedthrough schematic.png|center|frame]]
 
[[File:Feedthrough schematic.png|center|frame]]
Line 68: Line 66:
  
 
This frequency is set by a 2250 Ohm resistor and a 10 pF capacitor, resulting in a frequency-pass value of O(MHz). This reduces a large portion of the noise on the base signals and allows the dark noise and multiple PE signals to be seen. This information is useful for setting up and calibrating the readout and DAQ systems later.  This resistor can be changed out to control the filter, but this will also affect the gain on the readout signals. A larger resistor value will increase the gain on the system and vice versa. This should be balanced according to the signals achievable when all light leaks have been eliminated.
 
This frequency is set by a 2250 Ohm resistor and a 10 pF capacitor, resulting in a frequency-pass value of O(MHz). This reduces a large portion of the noise on the base signals and allows the dark noise and multiple PE signals to be seen. This information is useful for setting up and calibrating the readout and DAQ systems later.  This resistor can be changed out to control the filter, but this will also affect the gain on the readout signals. A larger resistor value will increase the gain on the system and vice versa. This should be balanced according to the signals achievable when all light leaks have been eliminated.
 
**Update 5/10/22**: Light leaks have been eliminated with a dedicated cleaning effort and vacuum-leak investigation. The noise which was previously present from the lamp fan has also been eliminated. Noise still remains from the amplifier - this has a regular shape so in theory could be subtracted from signals during analysis but we are working on reducing this further.
 
 
The amplifier is currently not producing any amplification of the signal. This is adjustable through the resistor values which still need to be balanced. This resistor is placed in a feedback position in order to control this amplification. However, the amplification currently acts on the current, which is later converted into a voltage signal. The Zurich group have sent us a spare board from their setup which utilizes a transimpedence amplifier. This works by converting the signal first into a voltage output and then amplifying it. A drawback is that it also flips the signal to a negative pulse. This should not be an issue as the digitiser ordered should be able to handle both positive and negative pulses. If it does matter, however, it is simple to flip the signal again to create positive pulses.
 
  
 
The feedthrough into the vacuum chamber is currently introducing a significant amount of noise into the system. To help to manage this, low-voltage feedthrough options are a possibility. The Carleton group have successfully used one such feedthrough with a reduced noise output. Several different low-voltage options are given below, though not all are appropriate for our electronics aims.
 
The feedthrough into the vacuum chamber is currently introducing a significant amount of noise into the system. To help to manage this, low-voltage feedthrough options are a possibility. The Carleton group have successfully used one such feedthrough with a reduced noise output. Several different low-voltage options are given below, though not all are appropriate for our electronics aims.
Line 94: Line 88:
 
| [[https://www.solidsealing.com/products/14/SMP-50-Ohm-Grounded-Shield-Recessed-Double-Ended-500V-CF2-75-Flange-Without-Plug/|FA33218-9FF-KF40]] || $815.00 || 9 || Up to 500 V || || Carleton group suggestion - Double Ended-Grounded Shield
 
| [[https://www.solidsealing.com/products/14/SMP-50-Ohm-Grounded-Shield-Recessed-Double-Ended-500V-CF2-75-Flange-Without-Plug/|FA33218-9FF-KF40]] || $815.00 || 9 || Up to 500 V || || Carleton group suggestion - Double Ended-Grounded Shield
 
|}
 
|}
 +
 +
[[Status update 5.10.22]]

Revision as of 14:51, 5 October 2022

XAMS SiPMs

Hamamatsu S13370-3050CN sensitive in the VUV region
Data sheet VUV SiPMs S13370 series
Data sheet (sensor specific)

XAMS filter electronics design shown below. This will be used to connect the sensor to a bias source as well as the amplifier to read out the signal.

Circuit design of signal filter boards used in XAMS lab.

Sensor troubleshooting

How to connect the sensor

The SiPM has a cathode and an anode as can be seen in the schematic drawing below, the drawing also shows how the SiPM is oriented as a diode. A SiPM needs to be reversed biased, which means the voltage at the cathode is higher than at the anode. This way there is no current flowing until the electric field is so high the SiPM brakes down, the brake down voltages of the used SiPMs is also shown below. Sensor 1804 was already marked as defect, but we still used it to test our electronics just to be sure it worked without breaking a new sensor. This sensor first gave signals which made it seem like it was still working, but after switching around electronics we were not able to get a signal. We therefore decided to check all the used sensors if they are broken. We decided upon the convention of using the wire with an extra bit of kapton for the red/cathode connection and the black/anode connection without any extra kapton.

Schematic drawing of the S13370-3050CN SiPM
Serial No. Vop at
M=2.55e6 [V]
Id at Vop [μA]
1804 56.58 0.46
1805 56.69 0.47
1822 56.51 0.50

Wanting to check if a sensor is working correct we asked Martin van Beuzekom for the R & D lab to help us and helped us with a method to determine if a sensor is broken. First we wanted to check if the anode and cathode were indeed oriented as was depicted on the schematics. To check this we need to characterize the voltage vs. current. The characteristic graph of how this would look like could be found below. So what we do to check, is connect the SiPM with what we think is the cathode to the positive of the voltage source and the anode to the negative. We then apply a very low (around 0.1 [V]) voltage on the SiPM and increase it in steps of 0.1 [V]. If the diode is broken the current will increase linearly, and when the voltage applied is negative it will be again linear. This implies the diode is just an open connection and thus broken. If when the positive voltage is applied we see an exponential growth in the current we know it is connected in forward biasing mode. We then flip the current and we then expect the current to increase only in the region of the breakdown voltage.
So sensor 1804 gave the linear characteristic, confirming to us it is broken and cannot be used to check our electronics.

Voltage vs. Current characteristic for a diode

Cable impedance and noise

This information about will be discussed in more detail on the page Impedance on cables.

Notes

RAW notes Martin on SiPMs and impedances

Grappa muon electronics

We were able to borrow a SiPM readout board from the muon project. The schematics of the electronics and refrence table can be found below.

CircuitMuon.PNG PCB refrence.PNG

SiPM readout installation

While the SiPMs will be operating inside the vacuum chamber, all electronics will be housed outside the chamber. This requires that the bias and signal be transmitted through a vacuum feedthrough on the side of the chamber. The readout system prepared for this case is shown below and detailed in this section. This only refers to use of the VUV sensitive SiPMs.

Amplifier board.png

In order for the SiPMs to operate, they require a bias voltage of 56 V (with the first signals visible at 53 V and no exceeding 57 V). We use a dedicated power supply to provide this bias. The bias supply and the signal cables must be fed through a feedthrough on the side of the chamber into the vacuum. The schematic created for the cabling through the three feedthroughs in place is shown below. This also includes the cables for the temperature readouts. This feedthrough has been creating a significant level of noise on the readouts. To reduce some of this noise, a Ferrite ring has been wound (twice) around the connecting cable. The connections of the cables to the inside of the feedthrough are currently individual cases and are very difficult to manually connect – a better solution to this will be considered.

Feedthrough schematic.png

The bias supply is passed through the circuit board shown above which also includes the amplifier and low-pass filter. The amplifier in the board also requires a power supply which is connected separately, but both are joined to a common ground, along with the circuit board and the vacuum chamber itself. The low-pass filter is used to eliminate any noise which has a higher frequency than the combination of resistor and capacitor will allow via the equation:

High-pass filter equation.png

This frequency is set by a 2250 Ohm resistor and a 10 pF capacitor, resulting in a frequency-pass value of O(MHz). This reduces a large portion of the noise on the base signals and allows the dark noise and multiple PE signals to be seen. This information is useful for setting up and calibrating the readout and DAQ systems later. This resistor can be changed out to control the filter, but this will also affect the gain on the readout signals. A larger resistor value will increase the gain on the system and vice versa. This should be balanced according to the signals achievable when all light leaks have been eliminated.

The feedthrough into the vacuum chamber is currently introducing a significant amount of noise into the system. To help to manage this, low-voltage feedthrough options are a possibility. The Carleton group have successfully used one such feedthrough with a reduced noise output. Several different low-voltage options are given below, though not all are appropriate for our electronics aims.

Model Quote Pins Electronic capacities Pressures Notes
[120XSD040-9] In use 9 Up to 500 V and 5 A per pin Down to 1e-8 mbar pressure
[18605-01-KF] €320.00 9 Up to 500 V and 5 A per pin 93.1 bar Single Ended-Grounded Shield
[21383-01-KF] €598.00 4 Single Ended-Grounded Shield
[DN40 CF-F-9 ] 9 Up to 1000 V and 8 A per pin Down to 1e-10 mbar pressure
[DN40 ISO-KF ] 7 Up to 380 V and 16 A per pin Down to 1e-8 mbar pressure Soldered on vacuum side
[[1]] $1,025.00 12 Up to 500 V Carleton group suggestion - Double Ended-Grounded Shield
[[2]] $815.00 9 Up to 500 V Carleton group suggestion - Double Ended-Grounded Shield

Status update 5.10.22