Dark count dependence on temperature

From Vacuum Ultra Violet
Jump to navigation Jump to search

The dark count rate depends on temperature. We use a peltier element and a dark box to quantify this relationship.

Setup

@Vikas: can you update this?

Measurements

We took dark count measurements at 12 different temperatures.

T = [9.565, 7.664, 1.915, 3.823, 11.484, 5.735, 17.256, 15.32, 13.405, 19.185, 21] degrees Celcius.

For each temperature, around 1500 waveforms were recorded with an oscilloscope. Each waveform is 5 microseconds. You can clearly see the single p.e. peaks already in the raw waveform shown in Figure 1. In Figure 2, a waveform with electronic noise is shown. For 9 degrees Celcius, we have only a few waveforms available.

Figure 1: Raw waveform at 13 degrees.

Analysis

For each temperature, we count the peaks in each recorded waveform. Because there is a lot of high-frequency noise on top of the signal, we low-pass filter the signals before peakfinding. The low-pass filter is a Butterworth filter with a cut-off frequency of 10MHz and order 1. The peakfinding uses a minimal height of 0.02, a width of 180, and a prominence of 0.05. The result is shown in Figure 3 for the signal in Figure 2. Even though there is noise, the filtering and peak finding steps manage to identify the correct signal peaks. In Figure 3, we see that this waveform counts 6 peaks. Figure 4 shows a histogram of counts for all 1500 waveforms for 1 degree Celcius. The distribution resembles a Poisson distribution fairly well. Therefore, we can use Poisson statistics to calculate the uncertainty. In Figure 5 you can see the resulting plot. Here we show the average number of peaks per waveform per temperature. The counts increase with temperature. We can convert these counts to a rate per second and fit the theoretical curve to the data. The final result is shown in Figure 6. We fitted Formula 1 to the measured points. From the best fit parameters we can obtain information about the average ionization energy Eg.

Formula 1: Dark count probability

Conclusions

Figure 2: Raw waveform at 13 degrees showing noise.
Figure 3: In blue the raw data, which is also in Figure 2. In orange the low-pass filtered data, and in red the peaks.
Figure 4: Histogram of the counts for all 1500 waveforms for 1 degree Celcius. In orange a Poisson distribution is shown with the same average as the blue histogram. The two distributions are similar.
Figure 5: The darkcounts per 5 μs with 1σ Poisson errors.
Figure 6: Dark count rate in MHz and temperature in Kelvin, with in orange a fit to the data.

We find Eg=0.7 eV. Given the silicon band gap energy of 1.1 eV this could indicate the presence of traps acting as Shockley-Read-Hall generation centers. However, since the doping of the SiPMs is currently unknown, it is hard to say anything about this.