Difference between revisions of "Spot size 23-11-2021"

From Vacuum Ultra Violet
Jump to navigation Jump to search
(Add plot of data)
m (Move plot)
Line 112: Line 112:
 
!|16.7 ± 0.2
 
!|16.7 ± 0.2
 
|}
 
|}
[[File:Beamsize.png|thumb|Beamsize measurement results]]
+
[[File:Beamsize.png|alt=|left|frameless|700x700px|Beamsize measurement results]]
 +
 
 +
 
 +
 
 +
 
 +
 
 +
 
 +
 
 +
 
  
  
  
 
[[Media:Measurment notes.pdf|Measurement notes]]
 
[[Media:Measurment notes.pdf|Measurement notes]]

Revision as of 09:19, 25 February 2022

On this page you will find the spot size measurments performed on 23 november 2021 by Emily, Marjolein and Casimir. Spotsizes were measured by tracing the beam shape on white paper in a cardboard holder placed inside the beam. Three different tracings were measured at each distance to get a more accurate measurement.

Measurement table

The distance in the table is measured to the inside of the vacuum chamber. To get the distance to the slit, 94 mm needs to be added to the distance in the table.

Measurements Entrance 3.05(mm) Exit 1.65 (mm)
Distance (mm) Measurer Horizontal (mm) Vertical (mm)
103 ± 1 Casimir 6.8 ± 0.2 9.1 ± 0.2
Emily 7.7 ± 0.2 9.5 ± 0.2
Marjolein 7.7 ± 0.2 9.5 ± 0.2
Average 7.4 ± 0.2 9.4 ± 0.2
126 ± 1 Casimir 7.5 ± 0.2 9.9 ± 0.2
Emily 6.9 ± 0.2 10.4 ± 0.2
Marjolein 6.5 ± 0.2 12.1 ± 0.2
Average 7.0 ± 0.2 10.9 ± 0.2
162 ± 1 Casimir 7.0 ± 0.2 12.3 ± 0.2
Emily 8.9 ± 0.2 12.7 ± 0.2
Marjolein 8.1 ± 0.2 12.0 ± 0.2
Average 8.0 ± 0.2 12.3 ± 0.2
175 ± 1 Casimir 9.1 ± 0.2 11.4 ± 0.2
Emily 8.8 ± 0.2 11.7 ± 0.2
Marjolein 6.8 ± 0.2 13.7 ± 0.2
Average 8.2 ± 0.2 12.3 ± 0.2
232 ± 1 Casimir 10.9 ± 0.2 18.0 ± 0.2
Emily 10.6 ± 0.2 15.9 ± 0.2
Marjolein 13.7 ± 0.2 14.9 ± 0.2
Average 11.7 ± 0.2 16.3 ± 0.2
251 ± 1 Casimir 10.8 ± 0.2 18.2 ± 0.2
Emily 11.6 ± 0.2 15.2 ± 0.2
Marjolein 12.8 ± 0.2 16.6 ± 0.2
Average 11.7 ± 0.2 16.7 ± 0.2






Measurement notes