Difference between revisions of "Calibrations"
Jump to navigation
Jump to search
Line 1: | Line 1: | ||
=====Calibration checklist===== | =====Calibration checklist===== | ||
− | * Clean chamber | + | * Clean chamber |
* Take apart | * Take apart | ||
− | * Clean modules | + | * Clean modules - BOSIMANITE AL-29 followed by ROGYPAL CA-307 Z, bake for 80*C for 18 hours |
− | * Clean O-rings | + | * Clean O-rings - Isopropanol |
− | * Put chamber together ( | + | * Put chamber together (check configuration required) |
− | |||
* Reconnect monochromator | * Reconnect monochromator | ||
* Adjust height to prevent bending of modules during pumping | * Adjust height to prevent bending of modules during pumping | ||
* Connect securely | * Connect securely | ||
− | * Clean sample holder | + | * Clean sample holder - isopropanol |
* Central piece | * Central piece | ||
* Three frames | * Three frames | ||
* Three doors | * Three doors | ||
* Suspension support | * Suspension support | ||
− | + | * Screws/springs | |
− | + | * Clean cooling apparatus - isopropanol | |
− | |||
− | * Clean cooling apparatus | ||
* Cooling braids | * Cooling braids | ||
* Screws/springs | * Screws/springs | ||
+ | * Clean base aluminium sample - isopropanol | ||
+ | * Clean readout equipment which will be inside chamber (wires) - isopropanol | ||
+ | * Clean SiPM holder - isopropanol | ||
+ | * Install temperature sensor | ||
+ | * Install sample holder with samples | ||
* Reconnect cooling apparatus | * Reconnect cooling apparatus | ||
* Cooling braids | * Cooling braids | ||
* Screws/springs | * Screws/springs | ||
− | |||
− | |||
* Install SiPM holder | * Install SiPM holder | ||
* Visually test that it is in line | * Visually test that it is in line |
Revision as of 14:25, 18 November 2022
Calibration checklist
* Clean chamber * Take apart * Clean modules - BOSIMANITE AL-29 followed by ROGYPAL CA-307 Z, bake for 80*C for 18 hours * Clean O-rings - Isopropanol * Put chamber together (check configuration required) * Reconnect monochromator * Adjust height to prevent bending of modules during pumping * Connect securely * Clean sample holder - isopropanol * Central piece * Three frames * Three doors * Suspension support * Screws/springs * Clean cooling apparatus - isopropanol * Cooling braids * Screws/springs * Clean base aluminium sample - isopropanol * Clean readout equipment which will be inside chamber (wires) - isopropanol * Clean SiPM holder - isopropanol * Install temperature sensor * Install sample holder with samples * Reconnect cooling apparatus * Cooling braids * Screws/springs * Install SiPM holder * Visually test that it is in line * Reconnect readout equipment * Test to make sure all connections are secure * Pull vacuum down to e-3mbar * Take a dark count run with aluminium sample and slits closed * 30s measurement time * Repeat in and out of the lamp beam if possible * With all SiPMs * Take temperature readouts * Measure lamp * Move grating with motor to calibrate motor control * Move grating and measure intensity at each wavelength (in 10nm steps) * Compare this to the expected spectrum * Repeat for all SiPMs * Compare spectra between SiPMs * Sweep SiPMs across the beam (in 5mm steps)