W0039_I11 50 um n-in-p
Depletion at -20 V
Breakdown at -160 V in the lab, but now it seems to be ~95V
Threshold at 700 e
Equalisation at -90 Volts

10. Small angle scan
Bias Voltage: -100 V
Threshold: 1000e
One spill per angle
Angles -21, -13 -8, -5, -3, -2, -1, 0, 1, 2, 3, 5, 8, 13, and 21

20. Bias scan for timing resolution
Telescope: 200 V
20 spills per step (~5 minutes per step)
Angle: 0 degrees
Bias voltages (negative!): 5, 10, 15, 20, 30, 40, 50, 60, 70, 80, and 90 V

21. Angle scan at grazing angles
Bias: -90 V, Telescope: 200 V
5 spills per step
Real angles: 86, 88, 88.3, 88.8, 89.1, 89.5, 90.3

22. Bias scan at grazing angle for timing resolution
Telescope: 200 V
20 spills per step (~5 minutes per step)
Angle: 89.5 degrees.
Bias voltages (negative): 15, 30, 60, and 90 V

23. 2D Scan: angle and bias
Telescope: 200 V
5 spills per step
Threshold: 800 e
Angles: 0 to 80 (inclusive) in steps of 5 degrees.
Bias voltages (negative): 15, 30, 60, and 90V

Angle [degrees], DUT X [mm]
0 47
5 46
10 46
15 45
20 44
25 44
30 43
35 42
40 42
45 41
50 41
55 40
60 40
65 40
70 39
75 39
80 39

30. Angle scan for spatial resolution
###REMEMBER TO FIRST CENTER THE DUT AT AN ANGLE 45 degrees###
10 spills per step
Bias voltage: 25 V (?), Telescope: 100 V
Angles: 38, 40, 42, 44, 45, 46, 48, 50, 52 degrees

50. Threshold scan for timing resolution
###CENTER dut FOR ZERO DEGREES###
20 spills per step
Bias voltage: -90 V, Telescope: 200 V
Angle: 0 degrees
Thresholds: 1600, 1400, 1200, 1000, 900, 800, 700, and lower if possible: 600, 500

60. Threshold scan for spatial resolution
###REMEMBER TO FIRST CENTER THE DUT AT AN ANGLE 45 degrees###
10 spills per step
Bias voltage: -25 V (?), Telescope: 100 V
Angle: 45 degrees
Thresholds: 1000, 900, 800, 700

Grazing angle measurements:

50. Threshold scan for timing resolution
###REMEMBER TO CHECK IF BEAM IS STILL CENTERED (MOVE TO DUT X 38 mm at grazing angles)###
10 spills per step
Bias voltage: -90 V
Angle: 89.5 degrees
Thresholds: 1600, 1400, 1200, 1000, 900, 800, 700, and lower if possible: 650, 600, 550, 500

40. Check for timing uniformity of DUT pixel matrix
###REMEMBER TO FIRST CENTER THE DUT AT AN ANGLE 0 degrees (It was DUT X 47 mm in the afternoon afternoon)###
For -90V bias, Telescope: 200 V
lowest thl possible (700?)
Angle: 0 degrees
20 spills per position (~5 minutes per step)
Depending on beam spot size, scan complete DUT (e.g. 1 around each corner and 1 in middle for big spot size)

41. Check for timing uniformity of DUT pixel matrix
For -90V bias, Telescope: 200 V
lowest thl possible
same as 40, but now 1 continues measurement while moving the DUT